“TF” SERIES

HIGH VOLTAGE TEST FIXTURE

  • Make accurate HV in-line measurements
  • View and measure AC ripple & noise on DC HV
  • Measure absolute HV DC to 0.25% @ 25 PPM stability
  • View and measure Trise, Tfall, overshoot & settling time
  • Measure & monitor signals from 35Hz to 10MHz
  • View signals from DC to 20MHz
  • View and measure AC mV on DC kV
  • PLC Analog/Digital Remote operation capability
  • GENERAL INFORMATION:
    The “TF” Series product line is designed to support the need to make accurate measurements of high-voltage power supply (HVPS) & HV system performance. These reduced-size HV test fixture devices can be used for research and development, incoming inspection, production testing, field testing, or calibration. Each “TF” Series device, when coupled with a conventional meter or oscilloscope, is a stand-alone test fixture optimized for a specific HV testing function. The “TF” Series is engineered to support accurate measurement of ripple, noise, pulses, absolute DC, DC stability, DC line regulation, DC load regulation, etc.

    KEY FEATURES:
    The UltraVolt “TF” Series models all feature dual Alden B110US10 HV connectors. These connectors facilitate in-line measurements as well as un-terminated measurements. Internal ARC limiting / softening resistors are present for safety. All “TF” models have the HV ground return connection isolated from the chassis ground connection by 100 kΩ and clamped by a protection device.

    SPECIFICATIONS:
    All specifications are subject to change without notice. UltraVolt will enhance specifications whenever possible, through continuous product and process improvement efforts. Customers are not contacted when changes are made unless they have arranged for configuration control with UltraVolt’s customer service department (“CSD”) through the “-Q” suffix program. Only the most significant items will be noted on UltraVolt’s web site, in the product change notice section.

    PACKAGE:
    The “TF” Series units are manufactured using the standard RF-tight chassis mount case. Overall size is 9.00” x 2.00” x 1.50” ±0.025”. Please see the Figure “G” outline drawing.

    ALTITUDE, HUMIDITY & TEMPERATURE:
    The “TF” Series operating performance is guaranteed between sea level and 10,000ft., in non-condensing relative humidity up to 95%, and between temperatures of -40°C to +65°C. Storage temperature range is -55°C to +105°C.

    PRICING & AVAILABILITY:
    The “TF” Series units are currently available with a stock to 3 weeks lead-time.


    “TF” SERIES MODELS:

    “Precision Divider”: 40TF-DCD
    A 40kV rated HV Test Fixture that features a precision 10,000:1 DC divider (“DCD”) with a full scale accuracy of ±1%, a temperature stability of better than ±25 ppm per °C, and a voltage coefficient of < 1% per 40,000 volts. DC Loading is 1 GigW. Capacitive loading is < 10pF.

    L-027

    WARNING! A shock hazard exists when the chassis ground or the HV return ground is not properly connected!


    L-026


    1800 OCEAN AVE., FRNT
    RONKONKOMA, NY 11779

    TEL 800-9HV-POWER
    TEL 631-471-4444
    FAX631-471-4696

    www.ultravolt.com

    next page

    © 2003-2008 UltraVolt, Inc.
    specifications subject to change without notice


    UltraVolt, Inc.
    1800 Ocean Ave. Frnt
    Ronkonkoma, NY 11779
    1-800-9HV-POWER
    1-800-948-7693
    631-471-4444
    Fax: 631-471-4696
    E-mail: csd@ultravolt.com